Clinical Validation of a Deep-Learning Segmentation Software in Head and Neck: An Early Analysis in a Developing Radiation Oncology Center
Andrea D’Aviero, Alessia Re, Francesco Catucci, Danila Piccari, Claudio Votta, Domenico Piro et al. · 2022 · 37 citationsRead the paper
BACKGROUND: Organs at risk (OARs) delineation is a crucial step of radiotherapy (RT) treatment planning workflow. Time-consuming and inter-observer variability are main issues in manual OAR delineation, mainly in the head and neck (H & N) district. Deep-learning based auto-segmentation is a promising strategy to improve OARs contouring in radiotherapy departments. A comparison of deep-learning-generated auto-contours (AC) with manual contours (MC) was performed by three expert radiation oncologists from a single center. METHODS: Planning computed tomography (CT) scans of patients undergoing RT treatments for H&N cancers were considered. CT scans were processed by Limbus Contour auto-segmentation software, a commercial deep-learning auto-segmentation based software to generate AC. H&N protocol was used to perform AC, with the structure set consisting of bilateral brachial plexus, brain, brainstem, bilateral cochlea, pharyngeal constrictors, eye globes, bilateral lens, mandible, optic chiasm, bilateral optic nerves, oral cavity, bilateral parotids, spinal cord, bilateral submandibular glands, lips and thyroid. Manual revision of OARs was performed according to international consensus guidelines. The AC and MC were compared using the Dice similarity coefficient (DSC) and 95% Hausdorff distance transform (DT). RESULTS: A total of 274 contours obtained by processing CT scans were included in the analysis. The highest values of DSC were obtained for the brain (DSC 1.00), left and right eye globes and the mandible (DSC 0.98). The structures with greater MC editing were optic chiasm, optic nerves and cochleae. CONCLUSIONS: In this preliminary analysis, deep-learning auto-segmentation seems to provide acceptable H&N OAR delineations. For less accurate organs, AC could be considered a starting point for review and manual adjustment. Our results suggest that AC could become a useful time-saving tool to optimize workload and resources in RT departments.
What the full text shows
Working prototype- Highest Dice similarity coefficient (DSC) achieved for the brain, eye globes, and mandible. — DSC 1.00 (brain), 0.98 (eye globes and mandible) (Limbus Contour software on H&N CT scans)
- Lowest overlap (DSC) observed for the cochleae and optic chiasm. — DSC 0.58 (left/right cochlea), 0.50 (optic chiasm) (Limbus Contour software on H&N CT scans)
- Highest 95% Hausdorff distance (HD) observed for pharyngeal constrictors and oral cavity. — 17.59 mm (pharyngeal constrictors), 12.67 mm (oral cavity) (Limbus Contour software on H&N CT scans)
- Lowest 95% Hausdorff distance (HD) observed for lenses and eye globes. — 0.57-0.75 mm (lenses), 1.03-1.13 mm (eye globes) (Limbus Contour software on H&N CT scans)
- Most structures analyzed showed high overlap with manual contours. — DSC > 0.85 for most structures (Limbus Contour software on H&N CT scans)
- Setup
- 12 patients with H&N cancers; 276 total contours analyzed across 23 OAR structures; CT scans acquired with GE RT discovery (0.625 mm slice thickness).
- How
- Planning CT scans were processed using Limbus Contour deep-learning software to generate auto-contours (AC). These were compared against manual contours (MC) created by expert radiation oncologists using Dice similarity coefficient (DSC) and 95% Hausdorff distance (HD).
- Limitations
- Small sample size (12 patients) · Single-center study · Larynx excluded due to guideline differences · Target structures excluded due to high delineation variability
Summarized by Gemma 4 from the open-access full text · 12 Nov 2026. Read it