Stress Relaxation Properties of Five Orthodontic Aligner Materials: A 14-Day In-Vitro Study

Paolo Albertini, Valentina Mazzanti, Francesco Mollica, Federica Pellitteri, Mario Palone, Luca Lombardo · 2022 · 38 citationsRead the paper

We aimed to investigate the stress relaxation properties of five different thermoplastic aligner materials subjected to 14 days of constant deflection. Five different thermoplastic aligner materials were selected, whose elastic properties varied: F22 Evoflex, F22 Aligner, Durasoft, Erkoloc-Pro and Duran. The static properties of these materials-in particular, stiffness, stress-strain curve and yield stress-were measured with a three-point bending test. For all the tests that were performed, a minimum of three samples per material were tested. The yield load, yield strength, deformation and particularly the stiffness of each material were found to be similar in the single-layer samples, while the double-layer samples showed far lower stiffness values and were similar one to another. F22 Evoflex and Erkoloc-Pro maintained the highest percentages of stress, 39.2% and 36.9%, respectively, during the 15-day period. Duran and Durasoft obtained the lowest final stress values, 0.5 MPa and 0.4 MPa, respectively, and the lowest percentage of normalized stress, 4.6% and 3.9%, respectively, during the 15-day period. All the materials that we tested showed a rapidity of stress decay during the first few hours of application, before reaching a plateau phase. The F22 Evoflex material showed the greatest level of final stress, with relatively constant stress release during the entire 15-day period. Further research after in vivo aging is necessary in order to study the real aligners' behavior during orthodontic treatment.

What the full text shows

Shown in the lab
  • F22 Evoflex maintained the highest final stress and the highest percentage of stress over 15 days. — 6.5 MPa final stress; 39.2% normalized stress (Constant deflection at 25% of yield strength in distilled water at 37 °C)
  • Erkoloc-Pro maintained the second highest percentage of stress over 15 days. — 36.9% normalized stress (Constant deflection at 25% of yield strength in distilled water at 37 °C)
  • Durasoft showed the lowest final stress and normalized stress over 15 days. — 0.4 MPa final stress; 3.9% normalized stress (Constant deflection at 25% of yield strength in distilled water at 37 °C)
  • Duran showed the second lowest final stress and normalized stress over 15 days. — 0.5 MPa final stress; 4.6% normalized stress (Constant deflection at 25% of yield strength in distilled water at 37 °C)
  • All tested materials exhibited rapid stress decay during the first few hours before reaching a plateau phase. (Constant deflection in distilled water at 37 °C)
  • Single-layer samples showed similar stiffness, yield load, and yield strength, while double-layer samples exhibited significantly lower stiffness. (Three-point bending test)
Setup
Five thermoplastic materials (F22 Evoflex, F22 Aligner, Durasoft, Erkoloc-Pro, Duran); minimum of 3 samples per material; samples immersed in distilled water at 37 °C.
How
Three-point bending tests were used to determine yield strength and stress relaxation. Long-term behavior (15 days) was predicted using the time-temperature superposition (TTS) principle by shifting 24-hour relaxation data collected at 47 °C to 37 °C.
Limitations
In-vitro study only; does not account for in-vivo aging or salivary enzymes. · Predicted long-term behavior relies on the validity of the TTS principle for these specific polymers. · Constant deflection does not simulate the intermittent loading of real clinical use.