Automatic 3-Dimensional Cephalometric Landmarking via Deep Learning

Gauthier Dot, Thomas Schouman, Shang‐Hung Chang, Frédéric Rafflenbeul, Adeline Kerbrat, Philippe Rouch et al. · 2022 · 70 citationsRead the paper

The increasing use of 3-dimensional (3D) imaging by orthodontists and maxillofacial surgeons to assess complex dentofacial deformities and plan orthognathic surgeries implies a critical need for 3D cephalometric analysis. Although promising methods were suggested to localize 3D landmarks automatically, concerns about robustness and generalizability restrain their clinical use. Consequently, highly trained operators remain needed to perform manual landmarking. In this retrospective diagnostic study, we aimed to train and evaluate a deep learning (DL) pipeline based on SpatialConfiguration-Net for automatic localization of 3D cephalometric landmarks on computed tomography (CT) scans. A retrospective sample of consecutive presurgical CT scans was randomly distributed between a training/validation set ( n = 160) and a test set ( n = 38). The reference data consisted of 33 landmarks, manually localized once by 1 operator( n = 178) or twice by 3 operators ( n = 20, test set only). After inference on the test set, 1 CT scan showed “very low” confidence level predictions; we excluded it from the overall analysis but still assessed and discussed the corresponding results. The model performance was evaluated by comparing the predictions with the reference data; the outcome set included localization accuracy, cephalometric measurements, and comparison to manual landmarking reproducibility. On the hold-out test set, the mean localization error was 1.0 ± 1.3 mm, while success detection rates for 2.0, 2.5, and 3.0 mm were 90.4%, 93.6%, and 95.4%, respectively. Mean errors were −0.3 ± 1.3° and −0.1 ± 0.7 mm for angular and linear measurements, respectively. When compared to manual reproducibility, the measurements were within the Bland–Altman 95% limits of agreement for 91.9% and 71.8% of skeletal and dentoalveolar variables, respectively. To conclude, while our DL method still requires improvement, it provided highly accurate 3D landmark localization on a challenging test set, with a reliability for skeletal evaluation on par with what clinicians obtain.

What the full text shows

Working prototype
  • Mean radial error for all landmarks on the test set — 1.0 ± 1.3 mm (Hold-out test set (n=37), excluding one outlier)
  • Success detection rates (SDR) for radial errors under 2.0, 2.5, and 3.0 mm — 90.4%, 93.6%, and 95.4% respectively (Hold-out test set (n=37), excluding one outlier)
  • Mean difference for angular and linear cephalometric measurements — -0.3 ± 1.3° and -0.1 ± 0.7 mm (Hold-out test set (n=37), excluding one outlier)
  • Measurements within Bland–Altman 95% limits of agreement for skeletal and dentoalveolar variables — 91.9% and 71.8% respectively (Comparison to manual landmarking reproducibility)
  • Inference time per CT scan — around 1 minute (Using the trained DL pipeline)
Setup
198 presurgical CT scans (128 training, 32 validation, 38 test) from patients with dentofacial deformities; 33 landmarks total (21 skeletal, 12 dental).
How
A coarse-to-fine deep learning pipeline based on SpatialConfiguration-Net (SCN) was implemented, using one network for downsampled full scans and five networks for full-resolution regions of interest. Performance was validated against manual annotations using mean radial error and success detection rates.
Limitations
One outlier case (cleidocranial dysplasia) showed very low confidence and high error for several landmarks · Does not automatically detect the presence or absence of landmarks (requires manual deletion of missing ones) · Limited to CT scans; not yet validated on CBCT

Summarized by Gemma 4 from the open-access full text · 7 Oct 2026. Read it