Seedlabs
Source research

Comparison of semi and fully automated artificial intelligence driven softwares and manual system for cephalometric analysis

Rumeesha Zaheer, Hafiza Zobia Shafique, Zahra Khalid, Rooma Shahid, Abdullah Jan, Tooba Zahoor et al. · 2024 · 13 citationsRead the paper

BACKGROUND: Cephalometric analysis has been used as one of the main tools for orthodontic diagnosis and treatment planning. The analysis can be performed manually on acetate tracing sheets, digitally by manual selection of landmarks or by recently introduced Artificial Intelligence (AI)-driven tools or softwares that automatically detect landmarks and analyze them. The use of AI-driven tools is expected to avoid errors and make it less time consuming with effective evaluation and high reproducibility. OBJECTIVE: To conduct intra- and inter-group comparisons of the accuracy and reliability of cephalometric tracing and evaluation done manually and with AI-driven tools that is WebCeph and CephX softwares. METHODS: Digital and manual tracing of lateral cephalometric radiographs of 54 patients was done. 18 cephalometric parameters were assessed on each radiograph by 3 methods, manual method and by using semi (WebCeph) and fully automatic softwares (Ceph X). Each parameter was assessed by two investigators using these three methods. SPSS was then used to assess the differences in values of cephalometric variables between investigators, between softwares, between human investigator means and software means. ICC and paired T test were used for intra-group comparisons while ANOVA and post-hoc were used for inter-group comparisons. RESULTS: Twelve out of eighteen variables had high intra-group correlation and significant ICC p-values, 5 variables had relatively lower values and only one variable (SNO) had significantly low ICC value. Fifteen out of eighteen variables had minimal detection error using fully-automatic method of cephalometric analysis. Only three variables had lowest detection error using semi-automatic method of cephalometric analysis. Inter-group comparison revealed significant difference between three methods for eight variables; Witts, NLA, SNGoGn, Y-Axis, Jaraback, SNO, MMA and McNamara to Point A. CONCLUSION: There is a lack of significant difference among manual, semiautomatic and fully automatic methods of cephalometric tracing and analysis in terms of the variables measured by these methods. The mean detection errors were the highest for manual analysis and lowest for fully automatic method. Hence the fully automatic AI software has the most reproducible and accurate results.

What the full text shows

Working prototype
  • Fully-automatic AI analysis (CephX) demonstrated the lowest detection error for the majority of parameters. — 15 out of 18 variables (Compared against manual and semi-automatic methods)
  • Semi-automatic AI analysis (WebCeph) demonstrated the lowest detection error for a small subset of parameters. — 3 out of 18 variables (Specifically Y-Axis, SNP, and MMA)
  • Inter-examiner reliability was highest for the fully-automatic method. — ICC 0.919 to 1.000 (Intra-group comparison between two investigators)
  • Inter-examiner reliability for the manual method was the lowest among the three. — ICC 0.503 to 0.953 (Intra-group comparison between two investigators)
  • Significant differences were found between the three methods for specific cephalometric variables. — 8 variables (Witts, NLA, SNGoGn, Y-Axis, Jaraback, SNO, MMA, and McNamara to Point A)
Setup
54 lateral cephalometric radiographs from patients (39% male, 61% female; mean age 16.7 years).
How
The study compared manual tracing (acetate sheets), semi-automatic AI (WebCeph), and fully-automatic AI (CephX) for 18 cephalometric parameters. Two investigators performed each method on the same set of radiographs to assess intra-group reliability (ICC, paired T-test) and inter-group accuracy (ANOVA, post-hoc).
Limitations
Potential errors in semi-automatic analysis of SNO (lowest ICC recorded) · Dependence on radiograph quality, density, and contrast for AI accuracy · Potential magnification errors in computerized software

Summarized by Gemma 4 from the open-access full text · 8 Oct 2026. Read it