Seedlabs
Source research

Accuracy of 3D-printed surgical guides compared with freehand technique for temporary anchorage device placement in the mandibular buccal shelf

Yasmin Youssef, Bayan Alyammahi, Amar Hassan Khamis, Abdel Rahman Tawfik, Ahmed A. M. Ghoneima · 2026 · 2 citationsRead the paper

Objective This in vitro study evaluated the accuracy of 3D-printed surgical guides for the placement of temporary anchorage devices (TADs) in the mandibular buccal shelf. Methods Fifty CBCT scans from the Dubai Dental Hospital database were used to create 100 3D-printed mandibular models. A total of 100 TADs (BENEfit “R” screw, 11 mm length, 2 mm diameter) were inserted in the buccal shelf region between the first and second molars. Each mandible was duplicated to form two groups: (1) a guided group, where TADs were placed using a custom-designed surgical guide, and a (2) manual group, where placement was performed freehand. Results The guided group showed significantly higher TAD placement accuracy than the manual group, with improved parallelism to roots, closer proximity to the buccal cortical bone, greater distance from the lingual cortical bone and root apices, more upright angulation, and increased safety from the inferior alveolar nerve ( p < 0.05). Placement success was 98% vs. 70% for guided and manual groups, respectively ( p < 0.001). Bone measurements showed greater width at the second molar and greater length at the first molar. Correlation analyses indicated that TAD angulation and proximity to cortical bone, root apices, and the inferior alveolar nerve were positively associated with local bone width, particularly at the second molar region ( p < 0.05). Conclusion The use of 3D-printed surgical guides significantly enhanced the accuracy, safety, and clinical reliability of TAD placement in the mandibular buccal shelf, offering a valuable adjunct to orthodontic practice.

What the full text shows

Shown in the lab
  • Higher TAD placement success rate for guided vs. manual placement. — 98% vs. 70% (Mandibular buccal shelf region)
  • Improved TAD angulation relative to the occlusal plane. — 76.90° ± 8.82 vs. 66.43° ± 11.05 (Guided vs. manual group)
  • Increased clearance from the inferior alveolar nerve (IAN). — 4.64 ± 1.89 mm vs. 3.32 ± 3.44 mm (Guided vs. manual group)
  • Greater distance from the lingual cortical plate. — 8.13 ± 2.48 mm vs. 6.67 ± 2.33 mm (Guided vs. manual group)
  • Closer proximity to the buccal cortical bone. — 4.11 ± 1.20 mm vs. 5.96 ± 1.54 mm (Guided vs. manual group)
  • Increased clearance from the distal root apex of the first molar. — 3.65 ± 1.67 mm vs. 2.17 ± 1.61 mm (Guided vs. manual group)
Setup
100 3D-printed mandibular models (derived from 50 CBCT scans), using BENEfit 'R' screws (11 mm length, 2 mm diameter).
How
CBCT scans were used to create 3D-printed mandibular models. TADs were inserted either freehand or using a custom 3D-printed surgical guide designed in Rhino 3D and printed with NextDent SG material.
Limitations
In vitro study using 3D-printed models rather than live patients. · All insertions performed by a single trained investigator. · Analysis limited to the right side of each model to minimize variability.

Summarized by Gemma 4 from the open-access full text · 1 Nov 2026. Read it